Empirical formulae for calculatingγ-ray detectors effective solid angleratio
Abstract
Determination of the detector effi ciency using volumetric cylindrical sources is very important in various
scientifi c and industrial fi elds, especially in the fi eld of quantitative analysis. To calculate the absolute activity
of any sample, the full-energy peak effi ciency (FEPE) of the detector is needed. By applying the effi ciency
transfer method, the FEPE of the detector would be determined easily without using the standard sources. This
approach depends on two main factors. The fi rst one, is the reference effi ciency of the reference source, which
is determined experimentally, and the second one, is the calculation of the effective solid angle ratio between
the sample and the reference source geometries. This work introduces an empirical formula for calculating
the second factor for using two different sizes of NaI(Tl) detectors. The validity of this empirical formula was
successfully demonstrated by comparing the calculating values with the experimental values
Author(s)
Mohamed Abd El Moneim
Coauthor(s)
Ahmed El Khatib, Mohamed El Zaher, Abouzeid Thabet, Mahmoud Abbas, Mona Gouda
Journal/Conference Information
Journal of Radiology and Oncology,Volume: 1, Issue: --, Pages Range: 012-021